|
Mass Spectrometry
Description
Mass spectrometry is an established analytical technique that identifies compounds by the mass
(more correctly, mass to charge ratio) of the analyte molecule. Mass spectrometry
is especially noteworthy among analytical techniques because the signals produced
by a spectrometer are the direct result of chemical reactions such as ionization
and fragmentation, rather than energy state changes that are typical of most
other spectroscopic techniques. Because of this distinction, mass spectrometry
is considered the only definitive analytical technique and detector.
The first general application of molecular mass spectrometry occurred in the early 1940s in the
petroleum industry for quantitative analysis of hydrocarbon mixtures in catalytic
crackers.
Coupling mass spectrometers with gas chromatographic systems allows separation and subsequent determination
of components of highly complex mixtures with a high degree of certainty.
Recently, manufacturers of mass spectrometers, particularly spectrometers coupled with gas chromatographs,
have significantly reduced their overall size and increased durability. This
allows what was once a laboratory bench top instrument to perform field analysis.
Typical Uses
In the environmental
field, mass spectrometers are typically used as detectors for gas chromatographs.
Because of the increased durability of modern instruments, field gas chromatographs/mass
spectrometers (GC/MS) are capable of the same analyses as fixed laboratory instruments.
EPA approved SW846 methods are then capable with field GC/MS
instruments. Some SW846 methods include:
- Semi-volatile
organic chemicals (SVOC) Method 8270C
- Base-neutral
acids Method 8270C
- Dioxins Method
8280
- MTBE Method
8260
- Volatile Organic
Chemicals (VOC) Method 8240
- Halogenated
VOC Method 8260B
Theory of Operation
As an analyte is introduced into the mass spectrometer, it is ionized by an
ion source, typically an electron stream. In the case of GC/MS, the analyte
eluting from the GC column is directed into the ion source of the mass spectrometer
through a heated interface. The loss of an electron by the analyte during ionization
generates the molecular ion. The charged molecular ion has the same molecular
weight as the analyte molecule.
An Electron beam of 20 electron volts (ev) is used to extract an electron from
a molecule. Excess energy from the beam will further fragment the molecular
ion to fragment (daughter) ions with lower mass to charge ratio (mass).
The positive ions produced by electron impact are attracted
through the slits of the ion source and mass analyzer. These ions are mass analyzed
for differentiation according to their mass-to-charge ratios.
The mass sorted ions are detected by an electron multiplier and the resulting
signal is sent to a data system for processing. A display of the electron multiplier
signal generated by the sorted molecular ions is displayed as the mass spectrum.
In the case of a coupled GC/MS, one type of data display is an integrated sum
of signals generated by the molecular ions plotted as a function of elution
time from the GC column. This representation is a total ions
current (TIC) chromatogram. Mass spectra can be extracted from each chromatographic
peak of the TIC.
The mass
spectrum is in the form of a bar graph that relates the relative intensity
of mass peaks to their mass-to-charge ratio. The largest peak in each spectrum
is termed the base peak. The heights of the remaining peaks are computed as
a percentage of the base peak height. The spectrum may be compared to a spectral
library for identification of the compound based on fragmentation pattern and
peak ratios.
The individual
TIC and ion peak intensities are directly related to the concentration of analyte
in the sample extract making GC/MS an extremely powerful analytical tool for
positive identification and quantitation of organic compounds.
System Components
Mass spectrometers consist of four basic components; ion sources, sample inlet systems, mass analyzers,
and a transducer. A brief description of various types of each component is
presented below.
Ion Sources
The starting point for a mass spectrometric analysis is the formation of gaseous analyte
ions. Ionization methods fall into two categories: gas phase sources and desorption
sources. Gas phase ionization relies on first vaporizing the analyte then ionizing
it. Desorption sources ionize the analyte in a solid or liquid state. Click
here for a list of ionization sources
Sample Inlet Systems
The purpose of the inlet system is to permit the introduction of a representative sample
into the ion source. Types of inlets include batch inlets, direct probe inlets,
and chromatographic inlets. Again, mass spectrometers used for environmental
analysis are essentially detectors for gas chromatographs. The flow rate from
capillary chromatographic columns is generally low enough that the column output
can be fed directly into the ionization chamber of the mass spectrometer. For
packed or magabore capillary columns, a jet separator is used to eliminate most
of the carrier gas from the analyte. The removal of carrier gas is critical
because the MS is operated under high vacuum. Introduction of carrier gas, therefore,
would disturb the high vacuum.
Mass Analyzers
The function of the mass analyzer is analogous to that of the grating in an optical spectrometer.
In mass spectrometers, dispersion is based upon the mass-to-charge ratios of
the analyte ions rather than upon the wavelength of photons. Mass spectrometers
are therefore categorized based on the type of mass analyzer in the instrument.
Two types of mass analyzers are typically used for GC/MS analysis: (1) the quadrupole mass analyzer, and (2) magnetic sector analyzer. The
quadrupole MS interfaced with capillary-column GC, is the most commonly used in an
environmental laboratory and field analysis. The sample extract is injected
onto the capillary column on the GC, where the individual compounds in the complex
mixture are separated. The individual compounds elute through the column at
different rates into the MS for detection.
The second type of mass analyzer typically used in mass spectrometers is the magnetic
sector. Magnetic sector mass spectrometers, also known as high resolution
mass spectrometers, are significantly more sensitive than the quadrupole. For
example, magnetic sector spectrometers detect dioxins at parts per trillion
(ppt) levels in soil and parts per quadrillion (ppq) in water. Typically, high
resolution MS is not used for field-based environmental analysis because of
size and stability requirements of the instrument.
Transducers
Like an optical spectrometer, a mass spectrometer contains a transducer that converts the beam
of ions into an electrical signal that can be processed, stored in computer
memory, and presented in graphical form. Transducers commercially available
for mass spectrometers include electron multipliers
and the Faraday Cup. The electron
multiplier is the transducer of choice for most spectrometers. Less common
detection systems include photographic plates and scintillation counters.
Mode of Operation
Mass spectrometers are operated under high vacuum to remove atmospheric interferences
that would affect the ionization process. All MS have safety devices so that
ion source cannot become operational until high vacuum is attained. Because
most MS are operated in conjunction with a gas chromatography, the removal of
carrier gas is critical because the MS is operated under high vacuum. Introduction
of carrier gas, therefore, would disturb the high vacuum.
Computers and microprocessors are integral in modern mass spectrometers. A characteristic of a mass spectrum
is the wealth of structural data that it provides. For structural determination
of a molecule, the heights and mass-to-charge ratios of each fragmentation peak
in a spectrum is determined, stored, and displayed. Because the amount of information
is so large, rapid acquisition and processing is essential. The microprocessor
is essential for mass spectrometric data acquisition and manipulation. Several
instrumental variables are controlled and monitored during data collection.
First, the computer serves as the instrument controller. Operating parameters
are set through communication via the keyboard. Secondly, the computer controls
programs responsible for data manipulation and output.
The digitized ion-current signal is processed prior to display. The peaks are normalized,
that is, the height of each peak relative to a reference peak is calculated.
The largest peak in the spectrum, the base peak, is arbitrarily assigned a peak
height of 100 (sometimes 1000). The mass to charge (m/z) value for each peak
must also be determined. This assignment is frequently based on the time of
the peak's appearance and the scan rate. Data are acquired as intensity versus
time during a controlled scan of the magnetic or electric fields. Conversion
from time to m/z requires careful periodic calibration. Calibration, or often
called tuning, is achieved with perfluorotri-n-butylamine (PFTBA).
For high resolution analysis, the standard may be added with the sample. The
computer is programmed to recognize and use the peaks of the standard as references
for mass assignments. For low resolution instruments, the calibration is obtained
separately from the sample due to the possibility of peak overlap.
Mass spectrometry has been widely applied to the quantitative determination of one or more components
of complex organic systems encountered in studies of environmental problems,
as well as the petroleum and pharmaceutical industries. Currently, analyses
are performed by passage of the sample through a chromatographic column and
into the spectrometer. With the spectrometer set at a suitable mass-to-charge
(m/z) ratio, the ion current is recorded as a function of time. This technique
is called selected ion monitoring. Generally, the areas are directly proportional
to the component concentrations. In this type of analysis, the mass spectrometer
simply serves as a sophisticated selective detector for quantitative analysis.
In another type of quantitative mass spectrometry for molecular species, analyte concentrations
are obtained directly from the heights of the mass spectral peaks. For simple
mixtures, it may be possible to find peaks at unique m/z values for each component.
Under these circumstances, calibration curves of peak heights versus concentration
are constructed and used for analysis of unknowns. More accurate results can
be achieved by incorporating a fixed amount of an internal standard substance
in both samples and calibration standards. The ratio of peak intensity of the
analyte species to that of the internal standard is then plotted as a function
of analyte concentration. The internal standard reduces uncertainties in sample
preparation and introduction.
Target Analytes
- Semi-volatile organic chemicals (SVOC)
- Base-neutral acids
- Dioxins
- Volatile Organic Chemicals (VOC)
- Halogenated VOC
- Pesticides and PCBs
Performance Specs
Performance specs include information on detection limits, calibration, and quality control.
Detection Limits
Part per billion detection limits have been reported for quadrupole mass spectrometers. Magnetic
sector mass spectrometers, also known as high resolution mass spectrometers,
are significantly more sensitive than the quadrupole. For example, magnetic
sector spectrometers detect dioxins at parts per trillion (ppt) levels in soil
and parts per quadrillion (ppq) in water.
Calibration
Analyses are performed by passage of the sample through a chromatographic column and into the spectrometer.
With the spectrometer set at a suitable mass-to-charge (m/z) ratio, the ion
current is recorded as a function of time. This technique is called selected
ion monitoring. Generally, the areas are directly proportional to the component
concentrations. In this type of analysis, the mass spectrometer simply serves
as a sophisticated selective detector for quantitative analysis.
In another type of quantitative mass spectrometry for molecular species, analyte concentrations
are obtained directly from the heights of the mass spectral peaks. For simple
mixtures, it may be possible to find peaks at unique m/z values for each component.
Under these circumstances, calibration curves of peak heights versus concentration
are constructed and used for analysis of unknowns. More accurate results can
be achieved by incorporating a fixed amount of an internal standard substance
in both samples and calibration standards. The ratio of peak intensity of the
analyte species to that of the internal standard is then plotted as a function
of analyte concentration. The internal standard reduces uncertainties in sample
preparation and introduction.
Quality Control
Method blanks are periodically analyzed to monitor laboratory and instrument-induced contaminants.
A method blank must not contain any analyte in a concentration higher than the
practical quantitation limit.
Matrix spike and
matrix spike duplicate samples are analyzed to evaluate the efficiency of the
sample preparation, precision of the analysis, and matrix effect.
Laboratory control
samples are used to evaluate the accuracy of the analysis. The control samples
are obtained from outside sources and contain known amounts of analytes. The
values obtained by analysis of the control samples are compared with the known
true values. The supplier of the control samples usually provides control limits.
The results obtained should fall within the published range of acceptance values.
Advantages
- Mass spectrometry
provides definitive compound identification.
- As a GC detector, identification of compounds that cannot be achieved by GC-PID, GC-FID, or
GC-ECD is possible.
- Field portable quadrupole mass spectrometers are now available.
Limitations
- Instrumentation is relatively expensive.
- Instrumentation operation requires a higher degree of expertise than most other instrumentation.
Cost Data
Mass spectrometer costs vary significantly. Instrument design and accessories affect instrument prices. Manufacturers listed below should be contacted directly for cost information.
Verification/Evaluation Reports
Verification of the performance of site characterization and field analytical technologies is
conducted through a variety of programs. Evaluation and verification reports
from EPA's Superfund Innovative Technologies Evaluation (SITE) Measuring and
Monitoring Program, EPA's Environmental Technology Verification Program (ETV)
program, along with links to certification statements from California EPA's
(CalEPA) California Environmental Technology Certification Program, are provided
below.
Superfund
Innovative Technologies Evaluation (SITE) Measuring and Monitoring Program
The SITE Demonstration Program encourages the development and implementation
of innovative treatment technologies for (1) remediation of hazardous waste
sites and (2) monitoring and measurement. In the SITE Demonstration Program,
the technology is field-tested on hazardous waste materials. Engineering and
cost data on the innovative technologies are gathered so that potential users
can assess the technology's applicability to a particular site. Data collected
during the field demonstration are used to assess the performance of the technology,
the potential need for pre- and post-treatment processing of the waste, applicable
types of wastes and waste matrices, potential operating problems, and approximate
capital and operating costs. The following reports from the measuring and monitoring
program are available for mass spectrometry:
No reports available for this technology
EPA's
Environmental Technology Verification (ETV) Program
EPA's Environmental
Technology Verification (ETV) Program verifies the performance of innovative
technologies. ETV was created to substantially accelerate the entrance of new
environmental technologies into the domestic and international marketplaces.
ETV verifies commercialized, private sector technologies. After the technology
has been tested, the companies will receive a verification report that they
can use in marketing their products. The results of the testing also are available
on the Internet. The following reports from the ETV program are available for
mass spectrometry:
- Bruker-franzen Analytical Systems, Inc Model EM640TM was verified for measurement of volatile
organics in soil, water, and soil gas. The verification documents available
consist of a verification report and
verification statement.
- Inficon, Inc. - HAPSITE with Headspace Sampling Accessory was verified for measurement of
chlorinated volatile organic compounds in water. The verification documents
available consist of a verification report
and verification statement.
- Viking SpectraTrakTM 672 was verified for measurement of volatile organics in soil, water, and
soil gas. The verification documents available consist of a verification
report and verification statement.
California
EPA's California Environmental Technology Certification Program
CalEPA's environmental technology certification program
is a voluntary program that provides participating technology developers, manufacturers,
and vendors an independent, recognized third-party evaluation of the performance
of new and mature environmental technologies. Developers and manufacturers define
quantitative performance claims for their technologies and provide supporting
documentation; CalEPA reviews that information and, when necessary, conducts
additional testing to verify the claims. The technologies, equipment, and products
that are proven to work as claimed are given official state certification. The
certification program is voluntary and self-supporting. Companies participating
in the program pay the costs of the evaluation and certification of their technologies.
Technologies
that have been certified through this program are listed below. Links are provided
to the web sites that provide the Certified Environmental Technology Transfer
Advisory and Certification Notice for the technologies.
No reports available for this technology.
Disclaimer
http://cluin.org/char/technologies/mspec.cfm
Page Last Modified: November 30, 2006

|